Cal Flame FM-204 Operator's Manual Page 88

  • Download
  • Add to my manuals
  • Print
  • Page
    / 309
  • Table of contents
  • TROUBLESHOOTING
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 87
78 Replacing the Hydrophobic Filter EAGLE 2 Operator’s Manual
6. Locate the hydrophobic filter. It is over the oxygen sensor. Note which
side of the hydrophobic filter has the RKI logo and part number. This is
the inlet side and should be facing toward the front of the EAGLE 2.
Figure 20: Replacing the Hydrophobic Filter
7. Pull the grey connecting tubes off of each end of the filter and remove it.
8. Install the new filter with the red RKI logo and part number on the inlet
side of the flow chamber, facing the front of the EAGLE 2. Make sure to
push the grey connection tubes all the way onto the filters hose barbs.
9. Confirm that the main PCB is seated in its slots and that its bottom edge
is resting on the bottom of the bottom case. If the main PCB is not seated
properly, then it may be damaged when the top case is re-installed.
10. Make sure that the top case gasket is fully seated in its groove and
carefully put the top case back on the bottom case. If you have any
difficulty mating the top and bottom cases, inspect the placement of the
main PCB and the placement of the top case gasket.
11. Turn the EAGLE 2 upside down and tighten the three case screws to
secure the top case to the bottom case.
Grey Connecting
Tube, 2X
Hydrophobic Filter
RKI Logo & Part Number on This Side
Bottom Case Shown Without
Tubing For Clarity
Page view 87
1 2 ... 83 84 85 86 87 88 89 90 91 92 93 ... 308 309

Comments to this Manuals

No comments