Cal Flame FM-204 Operator's Manual Page 33

  • Download
  • Add to my manuals
  • Print
  • Page
    / 309
  • Table of contents
  • TROUBLESHOOTING
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 32
EAGLE 2 Operator’s Manual Start Up 23
4. The LCD will show the following screen for about ten seconds.
5. The Battery Voltage Screen appears for a few seconds.
6. The Active Gases Screen appears for a few seconds indicating which
channels are active and their target gas.
7. If LUNCH BREAK is turned on (see “Updating the Lunch Break
Setting” on page 129), the Resume Measurement Screen appears. The
unit counts down from 5 seconds in the lower right corner of the LCD to
the right of “MEASUREMENTS”.
To continue accumulating peak and time-weighted average (TWA)
readings from the last time the EAGLE 2 was used, press and release
the AIR YES button before the countdown reaches 0 or allow the
countdown to reach 0. If you do not press the AIR YES button
within the 5 second countdown, the EAGLE 2 automatically resumes
accumulating the peak and TWA readings.The EAGLE 2 will also
continue to keep track of operating time including the operating time
GAS MONITOR
EAGLE 2
BATTERY MIN:
4.3 VOLTS
BATTERY NOW:
5.2 VOLTS
ACTIVE GASES
CH4 OXY
H2S CO
LUNCH BREAK MODE ON
RESUME
PEAK AND TWA
MEASUREMENTS? 2
Page view 32