112 Using Setup Mode EAGLE 2 Operator’s Manual
NOTE: If you define a gas whose LEL is above 50,000 ppm, the %LEL
reading in Measuring Mode will reflect the defined ppm ratio, but
the ppm reading in Measuring Mode will not indicate above 50,000
ppm. For example, if you set the ratio to be 150,000 ppm and set the
catalytic combustible channel to display the reading in ppm, the gas
reading will not indicate higher than 50,000 ppm, the equivalent of
33 %LEL and 5% volume for this ratio, but will continue to indicate
%LEL readings up to 100 %LEL and %volume readings up to 15
%volume, the equivalent of 150,000 ppm, if the display units are
changed to %LEL or %volume. In addition, all adjustable ppm
parameters cannot be set higher than 50,000 ppm.
10. Press and release POWER ENTER RESET to enter the ppm ratio. The
FACTOR for the gas begins to flash. The FACTOR for the gas is the
response factor for the user defined gas relative to methane. The response
factor must be obtained by testing the user defined gas and comparing its
response to methane. This parameter is used by the relative response
feature. See “Catalytic Sensor Relative Response Screen” on page 43 for
a description of the relative response feature and how to use it. See
“Combustible Gas Detection” on page 31 for a list of response factors for
several common hydrocarbon gases that have already been tested.
11. Use AIR YES and RANGE SHIFT to increase or decrease the
response factor to the desired number.
12. Press and release POWER ENTER RESET to enter the response factor.
The sensor voltage setting screen appears with the sensor voltage
flashing. The current gas and range is shown at the top of the screen and
the current sensor voltage is shown at the bottom of the screen.
The sensor voltage setting defines whether the catalytic sensor voltage is
set for full response, 2.40 volts, or methane elimination, 1.30 volts. If the
sensor voltage is set to 2.40 volts, the unit will default to Full Response
Mode when turned on, but the methane elimination feature can be turned
on in the Methane Elimination Mode Screen in Display Mode. If the
sensor voltage is set to 1.3 volts, the unit will default to Methane
Elimination Mode when turned on, but the methane elimination feature
can be turned off in the Methane Elimination Mode Screen in Display
CAT(0-50000ppm CH4)
SETTING EV
UP/DOWN THEN ENTER
2.40