Cal Flame FM-204 Operator's Manual Page 153

  • Download
  • Add to my manuals
  • Print
  • Page
    / 309
  • Table of contents
  • TROUBLESHOOTING
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 152
EAGLE 2 Operator’s Manual PID Calibration 143
PID Calibration
A PID channel can be calibrated using the auto calibration method or the
single calibration method. To calibrate a PID channel using the single
calibration method, see “Calibrating Using the Single Calibration Method”
on page 64 and follow the instructions for calibrating a single channel.
NOTE: The factory setting for the Span Factor menu item in Setup Mode is
OFF for an EAGLE 2 with a PID channel. So the range of
adjustment is not displayed when a single calibration is performed
for any channel if a PID sensor is installed.
If your instrument is a multi-channel instrument that includes a high range
PID channel, RKI Instruments, Inc. recommends using the auto calibration
method for convenience. The calibration instructions below show a 5 channel
instrument which has the four standard channels, LEL/oxygen/H
2
S/CO, and
a high range PID channel calibrated to and setup for isobutylene (IBL). There
are two ways to calibrate an EAGLE 2 with a PID channel installed. The first
way, described below in “Calibrating with a 4-Gas Cylinder and a PID
Cylinder” on page 143, uses a 4-gas cylinder and an appropriate PID
cylinder. The second way, described below in “Calibrating with a 5-Gas
Cylinder” on page 152, uses a multi-gas cylinder that contains the target gas
for all installed sensors. Follow the set of instructions appropriate for your
cylinder configuration.
Calibrating with a 4-Gas Cylinder and a PID Cylinder
1. See “Calibration Supplies and Equipment” on page 56 for a description
of the necessary calibration supplies. In addition to an appropriate multi-
gas cylinder that is used to calibrate any active standard channels, you
will also need a cylinder to calibrate the PID channel. See Table 17 on
page 178 for available cylinders. Make sure your calibration cylinder is
appropriate for the PID detection range.
2. Find a fresh-air environment. This is an environment free of toxic or
combustible gases and of normal oxygen content (20.9%).
3. While in Measuring Mode, press and hold the RANGE SHIFT button,
then press the DISPLAY ADJUST NO button and release both buttons.
4. If the unit prompts you for the password, enter it by using the AIR
YES and RANGE SHIFT buttons to select each password number and
then pressing and releasing POWER ENTER RESET to enter the number
and move on to the next one.
Page view 152
1 2 ... 148 149 150 151 152 153 154 155 156 157 158 ... 308 309

Comments to this Manuals

No comments